The core component of an ultrasonic transducer is piezoelectric ceramics, and it can be said that the quality of various transducers and their different application scenarios are closely tied to the key parameters of piezoelectric ceramics. So, how can one understand the key parameters of piezoelectric ceramics? And how should different ultrasonic transducers select piezoelectric ceramics based on these key parameters?
The core parameters of piezoelectric ceramics are divided into five categories, and now we will explain them in an easy-to-understand manner:
1. Piezoelectric constant
d₃₃: This is the most critical parameter, measuring the efficiency of "electrical energy → mechanical energy" conversion. Simply put, it means "how much voltage is applied, how much the material deforms" (inverse piezoelectric effect), or "how much force is applied, how much charge is generated" (direct piezoelectric effect). A higher d₃₃ value indicates stronger electromechanical conversion capability and greater device sensitivity. Typical PZT ceramics generally range from 200-600 pC/N, while high-performance single crystals can reach the 2000 pC/N level.
g₃₃ (piezoelectric voltage constant): Measures the conversion efficiency of "mechanical energy → electrical energy." A higher g₃₃ produces stronger voltage signals under the same force, making it suitable for high-sensitivity sensors.
2. Electromechanical coupling coefficient k: Reflects the efficiency of mutual conversion between electrical and mechanical energy, being a dimensionless number ranging from 0 to 1. A higher value indicates more complete energy conversion, such as the planar coupling coefficient Kp of PZT ceramics, which can exceed 0.6. Different symbols (e.g., K₃₁, K₃₃, Kp, Kt) represent distinct vibration modes.
3. Mechanical Quality Factor Qm: Indicates the magnitude of energy loss during material vibration. A higher Qm means less loss and "cleaner" vibration, making it suitable for devices requiring sustained stable vibration, such as ultrasonic transducers and filters; however, excessively high Qm often leads to reduced piezoelectric performance.
4. Curie temperature Tc: Reflects the "heat resistance" limit. Beyond this critical temperature, the material loses its piezoelectricity (depolarization), and this process is generally irreversible. For ordinary PZT ceramics, it is approximately 150-350°C, making this parameter crucial for high-temperature applications.
5. Dielectric constant ε: Reflects the material's ability to store electric charge and also affects the capacitance and impedance matching of devices. Generally, a higher dielectric constant results in greater capacitance for fabricated devices, requiring attention to impedance matching in high-frequency applications. Dielectric constant (ε) and dielectric loss (tan δ): Reflect "capacitance" and "heat generation.".
These parameters are not independent; they often trade off against each other. For example, increasing the piezoelectric constant d₃₃ typically comes at the expense of Curie temperature or mechanical quality factor Qm. Conversely, hard doping to boost Qm may reduce d₃₃. Therefore, when selecting materials, one cannot simply prioritize a single parameter like "higher is better." Instead, trade-offs must be made based on specific application scenarios—such as requiring high Qm for ultrasonic cleaning, high d₃₃ for sensors, or ensuring Curie temperature in high-temperature environments.